Adult intensive care is the core setting, with the full mode range from controlled ventilation through to pressure-supported spontaneous breathing available on one unit.
Weaning is supported by SIMV with assisted spontaneous breathing, mandatory minute volume adjusting to what the patient achieves, and CPAP with pressure support, alongside a rapid shallow breathing index for assessment.
Patients making early spontaneous effort are accommodated by an open breathing system and a multi-sense trigger, with up to 180 L/min available to meet a demanding inspiratory effort.
Lung protective strategies are supported by pressure-limited ventilation, BIPAP and, where AutoFlow is enabled, delivery of a set tidal volume at the lowest pressure the lung requires at that moment.
Paediatric intensive care runs on the same platform from 3 kg, with tidal volumes from 20 mL and inspiratory flow scaled to 6 to 30 L/min.
Neonatal work is covered where the NeoFlow option is fitted, from 0.5 kg with tidal volumes from 3 mL and rates to 150 per minute.
Step-down and non-invasive support is available where the mask ventilation option is fitted, with leakage compensation correcting for what the interface contributes.